Skip to article frontmatterSkip to article content
Site not loading correctly?

This may be due to an incorrect BASE_URL configuration. See the MyST Documentation for reference.

Experiment 10 — The Balmer Series and the Rydberg Constant

Objectives

By the end of this experiment you should be able to:

Textbook connection

Read §10.1–10.5. In the nonrelativistic Coulomb model, the Bohr and Schrödinger treatments both give approximately

En=−13.6 eVn2,E_n = -\frac{13.6\ \text{eV}}{n^2} ,

and the transitions among these levels are the spectrum you will measure. The Balmer series — transitions down to n=2n = 2 — is the part that happens to fall in the visible, which is why a formula fitted to it in 1885 preceded any theory of atomic structure by nearly thirty years.

Theory

The Rydberg formula

Transitions from level nin_i to level nfn_f emit photons with

1λ=RH(1nf2−1ni2).\frac{1}{\lambda} = R_{\text{H}}\left(\frac{1}{n_f^2} - \frac{1}{n_i^2}\right) .

For the Balmer series nf=2n_f = 2, and the visible lines are

Representative observed Balmer line centers in standard air (NIST compilation, Table 10); fine-structure components and their blends can shift a reported center

Line

Transition

Observed λ\lambda (air, nm)

Color

Hα\alpha

3→23 \to 2

656.279

deep red

Hβ\beta

4→24 \to 2

486.135

cyan

Hγ\gamma

5→25 \to 2

434.047

blue-violet

Hδ\delta

6→26 \to 2

410.174

violet, faint

Rearranged, (10.2) says that a plot of 1/λvac1/\lambda_{\rm vac} against (1/4−1/ni2)(1/4 - 1/n_i^2) is a straight line through the origin with slope RHR_{\text{H}}. That is the fit you will do, using every measurable line at once rather than averaging separate estimates. It is a leading-order model: the tabulated observed centers need not agree with its predictions to their last quoted decimal place.

Hydrogen energy levels from n=1 to the continuum, with the four Balmer transitions from n=3,4,5,6 down to n=2 marked in color, and one representative Lyman and Paschen transition shown in gray for context.

Figure 10.1:The Balmer series is one column in a larger structure: every visible line you measure is a transition landing on n=2n=2; Lyman lines land on n=1n=1 (in the UV) and Paschen lines on n=3n=3 (in the IR) — neither visible to the eye, which is why Balmer’s formula came first.

Which Rydberg constant?

Two constants appear in the literature and they are not the same:

R∞=mee48ε02h3c=1.0973731568×107 m−1R_\infty = \frac{m_e e^4}{8\varepsilon_0^2 h^3 c} = 1.0973731568\times10^{7}\ \text{m}^{-1}

is the value for an infinitely heavy nucleus, and

RH=R∞1+me/Mp=1.0967758×107 m−1R_{\text{H}} = \frac{R_\infty}{1 + m_e/M_p} = 1.0967758\times10^{7}\ \text{m}^{-1}

is the value for hydrogen, corrected for the finite proton mass through the reduced mass μ=meMp/(me+Mp)\mu = m_eM_p/(m_e + M_p). They differ by about 0.054%0.054\%. Separating them requires an uncertainty budget smaller than that difference, including calibration, air conversion, and the limits of the simple line-center model. A small fit error alone does not establish a measurement of the proton-to-electron mass ratio. State whether your total uncertainty can distinguish the two constants.

Air versus vacuum

Wavelengths measured in air are shorter than in vacuum by the refractive index of air, n≈1.00027n \approx 1.00027 in the visible — a 0.027%0.027\% effect, comparable to the reduced-mass correction above. Tabulated “air wavelengths” (the values in the table above) already include it. Be explicit about which convention you are using, or you will chase a systematic of exactly the size of the physics you are trying to see. The refractive index varies with wavelength and ambient conditions. Use a wavelength-dependent standard-air conversion for a precise comparison, or include the approximation in your uncertainty budget.

Pre-lab

Apparatus

Light from a hydrogen discharge tube passes through an adjustable entrance slit and a collimating lens, then reflects from a grating. Four Balmer colors return on the incident side at different angles toward a movable viewer.

Figure 10.2:Reflective-grating spectrometer, shown schematically. Collimated light from the entrance slit reaches the grating; each Balmer line returns at a different angle on the incident side. Use the actual instrument geometry and its angle zero for conversion to wavelength.

Procedure

Before you calibrate

Part A — Calibration

  1. Set up the spectrometer and focus it on the entrance slit. Narrow the slit until the lines are sharp; note that narrowing further past this point costs intensity without improving resolution.

  2. With the mercury lamp, measure the positions of several resolved known lines across the available span. Resolve the close 577/579 nm pair before counting them separately. Record the line-center rule and repeats.

  3. Calibrate with the geometry of your instrument. For a normal-incidence transmission grating, dsin⁡θ=mλd\sin\theta=m\lambda. A reflective grating at oblique incidence needs both incidence and diffraction angles with a stated signed-angle convention; the normal-incidence formula does not apply to it. For a USB spectrometer, fit a justified low-order pixel-to- wavelength relation. Check the actual lamp or detector wavelength range.

  4. Check the fit and its prediction uncertainty. Plot signed residuals versus wavelength. Estimate line-center repeatability, uncertainty in the fitted calibration, and any model mismatch separately. Residual RMS alone is not a systematic uncertainty for each subsequent line.

  5. Use the helium line near 667.8 nm to extend calibration beyond Hα\alpha or reserve it as an independent extrapolation check. Do not use the same line for both roles. Other held-out lines can check the blue range.

[ ] Checkpoint 1. Show the instructor your calibration fit, residuals, and a held-out check. Resolve any gross misidentification or drift before measuring hydrogen; document the achieved calibration accuracy.

Part B — The Balmer lines

  1. Replace the calibration lamp with the hydrogen tube using the approved shutdown procedure. Recheck slit illumination and an accessible reference line after the swap; if the slit, grating, or fiber moved, recalibrate.

  2. Locate Hα\alpha through Hδ\delta where resolved and measurable. Measure each usable line at least five times. On a vernier instrument, approach from alternating directions to expose backlash.

  3. Hδ\delta is faint. If you widen the slit, repeat nearby blue calibration lines at that width and record any center shift or blending. Do not invent a center for an unresolved line; three well-measured lines can still fit the one-parameter model.

  4. Watch for molecular hydrogen bands — the tube emits from H2_2 as well as H, creating extra spectral features that are not the Balmer series. The Balmer lines are sharp and bright; the molecular bands are broad and clustered. Note in your notebook which features you rejected and why.

Part C — Re-calibration

  1. Return to calibration lamps and re-measure at least one blue and one red line, preferably bracketing the measured span. Treat a shift as evidence of drift and include its effect in the uncertainty budget.

Analysis

Wavelengths

Apply the calibration to each measured position. Combine repeat scatter with calibration prediction uncertainty, checks for drift and model mismatch, and the air-index conversion. A common calibration error correlates the reported wavelengths. Quote each λ\lambda with an uncertainty, and compare each with the observed line centers individually before fitting. At the few-pm level, include the difference between observed blended centers and the simple Rydberg model in the interpretation.

The Rydberg constant

import numpy as np
from scipy.optimize import curve_fit

ni = np.array([...])                  # upper n for each usable line, >= 3 lines
lam = np.array([...])                 # measured VACUUM wavelengths, meters
Clam = np.array([...])                # full wavelength covariance, m^2

x   = 0.25 - 1.0 / ni**2
y   = 1.0 / lam
Cy  = Clam / np.outer(lam**2, lam**2)  # propagate 1/lambda

popt, pcov = curve_fit(lambda x, R: R * x, x, y,
                       sigma=Cy, absolute_sigma=True)
R, sR = popt[0], np.sqrt(pcov[0, 0])
print(f"R_H = {R:.7g} +/- {sR:.2g} 1/m")

Convert measured air wavelengths to vacuum first (λvac=nairλair\lambda_{\text{vac}} = n_{\text{air}}\lambda_{\text{air}}). Use a wavelength-dependent air index at the recorded conditions if your precision requires it. The RHR_{\text{H}} model uses vacuum wavelengths. Build Clam from repeat variability and calibration-parameter covariance; include shared terms off the diagonal. Test additional plausible drift or line-center models separately if their effects cannot be represented by this matrix.

Fit with the intercept forced to zero, then use a free-intercept fit as a diagnostic. A nonzero intercept can arise from calibration, blending, or model limitations; it is not by itself a unique diagnosis.

The reduced-mass question

Only if the total uncertainty and model checks support distinguishing RHR_{\text{H}} from R∞R_\infty, explore

meMp=R∞RH,measured−1\frac{m_e}{M_p} = \frac{R_\infty}{R_{\text{H,measured}}} - 1

and compare with 1/1836.151/1836.15. Propagate the uncertainty in your measured RHR_{\text{H}} and state the assumptions behind this leading-order relation. Otherwise report the bound your data place on the difference and explain why they do not determine the mass ratio.

Post-lab questions

Going further